User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: PM 2517 Calibration (Ver1.0) DATE: 03-Oct-93 AUTHOR: User Contributed REVISION: 1.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 34 NUMBER OF LINES: 91 CONFIGURATION: Fluke 5700A ============================================================================= * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Source: Service Manual PM P/N: @ * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * TUR Configuration: * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * When the T.U.R. is < 2:1, the source instrument, e.g. 5700A, must be * characterized and appropriate corrections entered in the correction file * (e.g. 5700.COR). Where T.U.R.'s are > 2:1 but < 4:1, the test tolerance * is reduced by root sum square (rss) techniques. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * This procedure is not executable in MET/CAL versions below ver. 2.00 * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Fluke make no warranty, expressed or implied, as to the fitness or * useability of this procedure. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Main Procedure TESTS: Auto Calibration * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * The following subprocedures are called from this main procedure: * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ R K 1.002 HEAD INITIAL PROCEDURE 1.003 DRAW 2 71 2 78 89 15 56 85 1.004 DISP TURN POWER ON[DRAW5,50,F] 1.005 DISP IF DISPLAY IS BLANK THEN CHECKUP BATTERY IN THE BACK 1.006 HEAD {DC VOLTAGE TEST} 1.007 DISP CONNECT TESTLEADS TO UUT V-mA-OHM AND COMMON TERMINALS 1.007 DISP SET UUT ON V-DC[DRAW30,50,F] 1.008 5700 1 0.9V 0.05/ 0.2% 2W 2.001 5700 1 -0.9V 0.05/ 0.2% 2W 3.001 5700 10 9V 0.05/ 0.2% 2W 4.001 5700 10 -9V 0.05/ 0.2% 2W 5.001 5700 100 90V 0.05/ 0.2% 2W 6.001 5700 100 -90V 0.05/ 0.2% 2W 7.001 5700 1000 900V 0.05/ 0.2% 2W 8.001 5700 1000 -900V 0.05/ 0.2% 2W 9.001 5700 * S 9.002 HEAD {AC VOLTAGE TEST} 9.003 DISP SET UUT ON V-AC[DRAW30,50,F] 9.004 5700 1 0.9V 0.1/ 0.5% 50H 2W 10.001 5700 1 0.9V 0.1/ 0.5% 1kH 2W 11.001 5700 1 0.9V 0.1/ 1% 10kH 2W 12.001 5700 1 0.9V 0.1/ 5% 20kH 2W 13.001 5700 10 9V 0.1/ 0.5% 50H 2W 14.001 5700 10 9V 0.1/ 0.5% 1kH 2W 15.001 5700 10 9V 0.1/ 1% 10kH 2W 16.001 5700 10 9V 0.1/ 5% 20kH 2W 17.001 5700 100 90V 0.1/ 0.5% 50H 2W 18.001 5700 100 90V 0.1/ 0.5% 1kH 2W 19.001 5700 100 90V 0.1/ 1% 10kH 2W 20.001 5700 100 90V 0.1/ 5% 20kH 2W 21.001 5700 1000 600V 0.1/ 0.5% 50H 2W 22.001 5700 1000 600V 0.1/ 0.5% 1kH 2W 23.001 5700 * S 23.002 HEAD {DIRECT CURRENT TEST} 23.003 DISP SET UUT ON A-DC[DRAW30,50,F] 23.004 5700 100 90mA 0.1/ 0.8% 2W 24.001 5700 100 -90mA 0.1/ 0.8% 2W 25.001 5700 * S 25.002 DISP DISCONNECT TESTLEAD FROM V-mA-OHM TERMINAL 25.002 DISP AND CONNECT IT WITH THE 10A TERMINAL 25.003 5700 10 2A 0.1/ 0.8% 2W 26.001 5700 10 -2A 0.1/ 0.8% 2W 27.001 5700 * S 27.002 HEAD {ALTERNATING CURRENT TEST} 27.003 DISP SET UUT ON A-AC[DRAW30,50,F] 27.004 5700 10 2A 0.1/ 0.8% 50H 2W 28.001 5700 * S 28.002 DISP DISCONNECT TESTLEAD FROM 10A TERMINAL 28.002 DISP AND CONNECT IT WITH THE V-mA-OHM TERMINAL 28.003 5700 100 90mA 0.1/ 0.8% 50H 2W 29.001 5700 * S 29.002 HEAD {RESISTANCE TEST} 29.003 DISP SET UUT ON OHM[DRAW30,50,F] 29.004 5700 1 0.19kZ 0.1/ 0.5% 2W 30.001 5700 10 1.9kZ 0.1/ 0.5% 2W 31.001 5700 100 19kZ 0.1/ 0.5% 2W 32.001 5700 1 0.19MZ 0.1/ 1% 2W 33.001 5700 10 1.9MZ 0.1/ 1% 2W 34.001 END